How to Calibrate a Pressure Transmitter — Step-by-Step Guide
How to Calibrate a Pressure Transmitter — Step-by-Step Guide
Pressure transmitter calibration ensures that the transmitter provides accurate readings of applied pressure by comparing it with a known reference standard. This is essential for process accuracy, safety, and reliability.
Required Equipment
- Pressure Calibrator / Hand Pump / Dead Weight Tester – for applying known pressure.
- Digital Reference Gauge – for precise measurement accuracy.
- 24 V DC Power Supply – to power the transmitter.
- Multimeter / Loop Calibrator – to measure 4–20 mA output signal.
- HART Communicator – for SMART transmitters.
- Tubing and Fittings – for proper connection and sealing.
Working Principle
A pressure transmitter converts the applied pressure into a proportional electrical signal (4–20 mA). Typically, 4 mA represents 0% (LRV) and 20 mA represents 100% (URV) of the calibrated pressure range.
Calibration Procedure
Step 1: Isolate & Prepare
- Isolate the transmitter from the process.
- Ensure it is depressurized and in safe condition/ ideal condition before starting.
Step 2: Connect Calibration Equipment
- Connect the pressure source to the transmitter pressure port.
- Connect 24V DC power supply to power terminals.
- Connect multimeter/loop calibrator to read the 4–20 mA output.
- If applicable, connect the HART communicator for SMART calibration.
Step 3: Apply Known Pressure (5-Point Test)
Apply pressure at 0%, 25%, 50%, 75%, and 100% of the transmitter range (both increasing and decreasing). Record the readings and calculate error.
Example (0–7 bar transmitter):
| Applied Pressure | Expected Output |
|---|---|
| 0 bar (0%) | 4.00 mA |
| 1 bar (25%) | 8.00 mA |
| 3 bar (50%) | 12.00 mA |
| 5 bar (75%) | 16.00 mA |
| 7 bar (100%) | 20.00 mA |
Step 4: Zero & Span Adjustment
Manual (Analog Transmitters):
- Apply 0% pressure and adjust Zero to 4.00 mA.
- Apply 100% pressure and adjust Span to 20.00 mA.
- Repeat until both are correct.
SMART (HART Transmitters):
- Connect HART communicator → Setup → Calibration → Trim Output / Sensor Trim.
- Apply 0% pressure → Perform Lower Trim.
- Apply 100% pressure → Perform Upper Trim.
Step 5: Verification
- Repeat the 5-point test after adjustment.
- Ensure readings are within ±0.1% tolerance or as per ISO standards.
- Document calibration results and attach reference instrument details.
💡 Pro Tips
- Vent air from the line before calibration (especially for liquid services).
- For Differential Pressure (DP) transmitters, ensure correct high and low port connections.
- Allow 5–10 minutes for signal stabilization before recording readings.
- Perform recalibration every 6–12 months as per QA schedule.
Comments
Post a Comment